PK5064
PK5064 complete Graphical Power Quality Analysis System includes PS5000 Graphical Power Quality Analyzer with extended memory, AC charger, Quick-Start manual, deluxe voltage leads, four eFX6000 flexible current probes, SD memory card, CAS3 hard-shell carrying case, Windows PSM-A Software, and 1-year deluxe warranty.
This system contains the essentials for almost any power quality study, with the graphical insight of a colored visual display. Other standard systems that include the PS5000 are the PK5064-PRO and the PK5064-PRO+.
Features
Measure and log all common power measurement types: V, A, kW, kVA, kVAR, kWH, PF, unbalance, Hz, THD, individual harmonics, logging of generated power separately from consumed power, etc.
Measure and log expanded power measurement types: individual harmonics, voltage and current unbalance, displacement power factor (DPF), pos & neg VAR, separate recording of positive, negative, & net Power.
Capture power quality events in multiple ways to view and record swells, dips (sags), inrush, high speed absolute and relative transients, of both voltage and current. All inputs are sampled continuously and simultaneously so nothing is missed.
Wide range of fundamental frequency modes for 50Hz, 60Hz, 400Hz, variable frequency systems, and DC to cover all your needs.
Able to evaluate all power systems: 3 phase Wye, 3 phase Delta (including 4-wire, open, and grounded), split-phase, single phase, quad single phase (using same source) , 3PT/3CT, 3PT/2CT, 2 current power, DC, quad DC (using same source).
The included Advanced PowerSight Manager (PSM-A) Windows software improves your efficiency and eliminates sources of error at each stage of the monitoring process:
- TestPlan Manager™ to organize and automate your tests and the data retrieval
- SureStart™ expert system to advise if the connections or settings seem correct or not, prior to starting
- DataSetup Wizard™ to set your triggering and logging parameters according to your application
- ReportWriter Wizard™ to automatically generate error-free reports.
Benefit from the safety rating of 1000V CAT III, 600V CAT IV and Bluetooth Wireless Communications (remote viewing/control of the display without the burden of wearing PPE).
The Extended Memory comes installed in the PS5000. Paired with the SD Memory Card, the analyzer has virtually unlimited logging capacity. The SD Memory Card also provides redundant storage to reduce the risk of data loss.
Extend the measurement range with the use of our industry-widest array of accessories and options:
- measure down to 0.002 amps AC (HA-GFD)
- measure from 5 to 2000 amps DC (to 600A with DC600, to 2000A with DC2000)
- direct connect and measure to 15kV (to 5kV with 5KVP, to 15kV with 15KVP)
- measure power-line noise to 100kHz (FAO)
First rate technical support, direct from the manufacturer to you!
Specifications
General Specs
Category | Notes |
---|---|
Size | PS5000 Analyzer: 3.88" (9.86 cm) W x 7.72" (19.61 cm) L x 1.58" (4.01 cm) Deep at bottom end; 2.14" (5.44 cm) Deep at top end |
Weight | PS5000 Analyzer: 1.1 lbs (0.5 kg) |
Operating Range | 32 - 122 degrees F (0 - 50 degrees C) Relative humidity to 70% (non-condensing) |
Power Requirement | 12V DC @ 500 mA |
Battery Life | 8-10 hours after 4 hour charge |
Battery Type | Li-ion |
Sampling Rate | 8 microseconds for Voltage and Current, continuous/simultaneous |
Safety | 600V CAT IV, 1000V CAT III |
SD Card Slot | Accepts SD Cards to 32GB (SDHC) |
Logging Period (Resolution) | 1 Second - 99 Minutes (set by user) |
Logging Duration (Length of session) | up to 2 years, depending on memory allocation |
Internal Data Retention | 8 years |
Other Features | Crest Factor, K Factor, Peak Demand Period, Peak Demand of Peak, Demand Period, Duty Cycle, On/Off Cycle %, Avg. On Time, Avg Off Time |
Voltage Measurement
Category | Notes |
---|---|
Direct Measurement Range | 1 - 1,000V (up to 15,000V with optional 15KVP) |
Measurement Range | 1 - 6 megavolts (using input ratios with PTs) |
Accuracy | 0.1% of reading ±0.3 Vrms |
Current Measurement
Category | Notes |
---|---|
Direct Measurement Range | 1 - 6,000A AC (down to 0.002A with optional HA-GFD, DC to 2,000A with optional DC2000, etc) |
Display Range | 0.001 - 6 megamps (using input ratios and CTs) |
Accuracy | ±1 % of reading +/-0.2A in low range, +/- 2.0A in high range |
Power Measurement
Category | Notes |
---|---|
Direct Measurement Range | 1 - 18 megawatts (assuming 1000V, 6000A, PF=1.00, 3-phase power) |
Display Range | from 0.002W (using HA-GFD) to 999 megamps (using input ratios and CTs) |
Accuracy | ±0.5 % of reading +/- the accuracy of the current probes |
Components
PS5000 Graphical Power Quality Analyzer with extended memory
AC Charger (specify 120 or 230 volt version)
Four eFX6000 6000 amp flexible current probes
Deluxe Voltage Leads with Jumbo Alligator Clips
1-Year Deluxe Warranty