HA-GFD – 5mA High Accuracy Ground Fault Detector

Our HA-GFD Current Probe is our most sensitive current probe, measuring down to 5 milliamps.  Ideal for measuring leakage current and ground faults.  Dual range allows measuring up to 200 amps in high accuracy.

$995.00

Our HA-GFD Current Probe is our most sensitive current probe, measuring down to 5 milliamps.  Ideal for measuring leakage current and ground faults.  Dual range allows measuring up to 200 amps in high accuracy.

Features

Our most sensitive current probe, accurately measuring down to 5 milliamps.
Ideal for measuring leakage current and ground faults.
Large jaws allow clamping around all three phases to net out the currents and find the leakage current.
Dual measurement range allows measuring up to 200 amps in high accuracy, so it is useful for all small to mid-size AC loads.

Specifications

Low Range Measurement Accuracy:

±3% from 0.005 to 0.009 amps

±2% from 0.010 to 0.049 amps

±1% from 0.050 to 1.000 amps

Hi Range Measurement Accuracy:

±2% from 5 to 8.9 amps

±1% from 9 to 200 amps

Inside Diameter:  4.5 inches
Cable Length:  2 meters (6 feet).

PLEASE NOTE:  The probes are used exclusively with the PowerSight® line of Energy Analyzers, manufactured by Summit Technology Inc.
eFX6000 probe small

eFX6000 Flex Current Probes

HA1000 probe small

HA-1000 Clamp Current Probe

HA100 probe small

HA-100 Clamp Current Probe

DC2000 current probe

DC2000 DC Current Probe

CASA To Go

CASA Accessory Carrying Case